Burchell, Thomas R. and Boubertakh, Redha and Mohiddin, Saidi and Miquel, Marc E. and Westwood, Mark A. and Mathur, Anthony and Davies, L. Ceri (2011) Adenosine Stress Perfusion Cardiac MRI: Improving Image Quality Using a 32-Channel Surface Coil. Open Journal of Medical Imaging, 01 (02). pp. 21-25. ISSN 2164-2788
OJMI20110200007_79670274.pdf - Published Version
Download (337kB)
Abstract
Purpose: Adenosine stress CMR is commonly used to assess myocardial ischaemia. Obtaining high quality images requires maximising signal to noise ratio (SNR) over a large double-oblique field of view (FOV) whilst minimising artefacts. A 32-channel surface coil may provide a higher SNR over a larger FOV compared to standard coils, possibly leading to improved image quality. Materials and Methods: 50 adenosine perfusion CMR scans were performed on a Philips Achieva CV 1.5T, with either a 5 or 32-channel coil (25 patients each) using standardised acquisition protocols. 3 short axis slices were acquired per cardiac cycle and the resulting cine images were scored by two blinded CMR specialists on a quality scale of 1 to 5. Phantom studies were performed using similar acquisition parameters and the SNR was calculated and compared across a range of acceleration factors. Results: The mean patient age was 62 ± 11 years and 50% of patients were male. The image quality scores were higher using the 32-channel coil (mean 3.8 ± 0.7 vs 3.2 ± 0.9 p = 0.002). The average phantom SNR was greater for the 32-element coil across the range of acceleration factors measured (103 vs 86 p = <0.001). Conclusions: The 32-channel coil produces significantly higher quality images and a higher SNR than the 5-channel coil in routine perfusion CMR.
Item Type: | Article |
---|---|
Subjects: | Souths Book > Medical Science |
Depositing User: | Unnamed user with email support@southsbook.com |
Date Deposited: | 29 Mar 2023 07:45 |
Last Modified: | 03 Sep 2024 05:45 |
URI: | http://research.europeanlibrarypress.com/id/eprint/411 |